ATCD-415A Thermocouple ADC (cold junction compensation, 3.125 kHz b/w) - 15ch at 12.5 ksps
- Downloads
- SPECIFICATIONS
- SUPPORT
- REQUEST A QUOTE
Key features
- 15 grounded or isolated thermocouple input channels
- Three PT50/PT100 channels and one built-in top-block sensor for junction compensation
- Supports multiple thermocouple types with digital cold junction compensation and programmable thermocouple type per channel
- Accuracy (0.5°C typical for K-type in -50 to 150°C range, 1°C typical outside this range)
- 16-bit simultaneous sampling with three configurable output streams on each channel
- User defined linearization and compensation for errors in the entire measurement chain
- Open thermocouple detection
Applications
Thermocouple temperature measurements
ADAU-2006 Ultra-Compact, Scalable and Reliable Data Acquisition
The ADAU-2006 is an Axon Family ADAU chassis with an integrated, isolated 50W PSU, slots for up to 6 ADAU user modules and a separate slot for an Overhead module. A wide input operating range,...
APCM-404A-1 IRIG-106 PCM Encoder (PMF Output)
The APCM-404A is used to encode data in an IRIG-106 Chapter 4 (Ch.4) PCM stream. Parameters from any ADAU module in the chassis can be placed in the Ch.4 PCM stream. Parameters from other system...
ASCD-424A Differential Ended ADC (5 kHz b/w) - 24ch at 20 ksps
The ASCD-424A is used to condition and digitize up to 24 analog differential ended channels. Single ended configurations are possible by enabling programmable pull-down resistors on any of the...
The ATCD-415A is used to condition and digitize up to fifteen analog thermocouple channels, three 4-wire RTD (dedicated for PT100) channels and offers top-block built-in temperature sensor.
The ATCD-415A performs linearization for the selected cold junction sensor and thermocouple, and compensates the thermocouple channel accordingly. Any of the three PT50/PT100 channels or top-block built-in temperature sensor can be selected as cold junction compensation.
The module offers programmable pull-down resistors for thermocouples isolated from the module ground.
At the heart of the ATCD-415A is a hard-wired state-machine that oversamples all channels at a rate between 50 ksps and 100 ksps and digitally filters any noise above the user-programmable cutoff frequency. This is achieved using cascaded, 15-tap FIR filters with output rate decimation, followed by a final FIR or IIR filter. If IIR filtering mode is selected, the last digital filter in the filtering chain is an 8th or 16th order (selectable) Butterworth filter. If FIR mode is selected, the last digital filter in the filtering chain is a 49-tap Kaiser window, Beta 6 filter.
There are three independently configurable output streams per channel, allowing different sample rate, cutoff, and filter type to be selected for each output stream.
All signals are sampled simultaneously. When several channels are sampled at different sampling rates, at the start of an acquisition cycle all channels are aligned.
Download Product Sheet register now
Product Finder
CONTACT SALES
Contact our sales team today to learn more about our products and services.
GET SUPPORT
Our support team can help answer your questions - contact us today.